TC-3 Temperature Controller User’s Manual · Page 37
Error Messages
Error
Code
Description Cause
E-
100
General Error The error code is non-specific, and is
generally used when no other error
code is suitable.
E-
102
Message too long The message is too long to process
(USB/Serial only).
E-
104
Type not allowed The RADix type was invalid
E-
123
Path not found The message used an invalid path
command (USB/Serial only).
E-
124
Data mismatch The message contained data that did
not match the expected format
(USB/Serial only).
E-
126
Too few or too
many elements
The command requires more or less
than the number of parameters
actually supplied.
E-
127
Change not
allowed
An attempt was made to change a
parameter that cannot be changed, or
is currently read-only.
E-
201
Data out of range The message attempted to set a value
that was outside the allowable range
(USB/Serial only).
E-
202
Invalid data type When trying to parse the message, the
data was in an invalid format
(USB/Serial only).
E-
204
Suffix not valid An invalid number base suffix (radix)
was encountered when parsing a
number (USB/Serial only).
E-
402
Sensor open,
output turned off
A sensor open circuit was detected and
the output was turned off.
E-
403
Module open,
output turned off
A Peltier module open circuit was
detected and the output was turned
off.
E-
404
I limit, output
turned off
A current limit was detected and the
output was turned off.
E-
405
V limit, output
turned off
A voltage limit was detected and the
output was turned off.
E-
406
Thermistor
resistance limit,
output turned off
The thermistor resistance limit (high or
low) was exceeded and the output was
turned off.
E-
407
Temperature limit,
output turned off
The temperature limit (high or low)
was exceeded and the output was
turned off.