Page 14 · TC-3 Temperature Controller User’s Manual
Menu Description Factory
Default
PID P The proportional term of the PID loop.
Will only be available if Gain is set to
PID.
1
PID I The integral term of the PID loop. Will
only be available if Gain is set to PID.
0.01
PID D The derivative term of the PID loop. Will
only be available if Gain is set to PID.
0
Sensor Sets the temperature measurement
sensor. Can be Therm 100uA, Therm
10uA, Therm, LM335, AD590, RTD, or
RTD (4-wire).
Therm
100uA
T-Low Lim T-Low Lim is the lower temperature
limit. If operating in T Mode and the
actual temperature drops below this
value, the output will be turned off.
Temperature limits not monitored in R
Mode or Ite Mode. Limit monitoring can
also be disabled in software (see
TEC:ENAB:OUTOFF command).
-99°C
T-High Lim T-High Lim is the upper temperature
limit. If operating in T Mode and the
actual temperature rises above this
value, the output will be turned off.
Temperature limits not monitored in R
Mode or Ite Mode. Limit monitoring can
also be disabled in software (see
TEC:ENAB:OUTOFF command).
125°C
R-Low Lim R-Low Lim is the sensor low limit. If
operating in R Mode and the actual
sensor measurement drops below this
value, the output will be turned off. R
limits are not monitored in Ite Mode.
Limit monitoring can also be disabled in
software (see TEC:ENAB:OUTOFF
command).
0.01kΩ