TC-3 Temperature Controller User’s Manual · Page 13
Settings and Menus
All parameters of the TC-3 can be viewed and changed within the
menu. The menus are constructed with the most used parameters first.
To change any setting, press Menu to enter the menu then rotate the
knob to select the parameter to change. Press the knob to begin
changing the value. As a visual indication that you are in edit mode,
you will see an asterisk appear next to the value. Once you have made
your change, press the knob or Menu button to store the value.
Pressing the Menu button will store and exit the menu, while pressing
the knob will store the value but leave you in the menu to make
additional changes.
Some settings are contained inside a sub menu, such as
communications settings. To access the sub menu, simply press the
knob to enter the sub menu when its name is displayed.
Below is a complete list of available settings:
Menu Description Factory
Default
Root Menu Main Menu
Mode This set the operating mode (T Mode, R
Mode, or Ite Mode) of the temperature
controller.
T Mode
Mount Specify the mount connected to the TC-
3. If using a LaserMount, select the
appropriate model. Otherwise, select
User Defined.
User
Defined
I Lim I Lim sets the current limit of the
temperature controller. The limit should
be set to a value that is suitable for
your Peltier device.
3A
Gain Gain controls the response of the
temperature controller. A higher gain
value will cause the controller to
respond more quickly to the difference
between the set point and the actual
temperature, while a lower value will
cause it to respond more slowly. Select
PID for direct access to the PID
parameters. Read more on setting gain
below.
30